Research Article | Volume: 9, Issue: 8, August, 2019

Validation of an HPLC-MS method for the determinatin of vardenafil in rat urine

Liudmila Osypchuk Iryna Halkevych Sophia Davydovych Yuriy Bidnychenko   

Open Access   

Published:  Aug 03, 2019

DOI: 10.7324/JAPS.2019.90811
Abstract

A simple and accurate technique of high-performance liquid chromatography with mass spectrometry was developed for the quantitative determination of vardenafil and its metabolites in urine. Extraction of vardenafil and internal standard (sildenafil) was performed from 5 ml of urine with 1,2-dichloroethane at pH = 7.5, followed by purification of samples on Oasis HLB cartridges. Chromatographic separation was carried out on a Zorbax SB-C18 reversed-phase column (50 mm × 2.1 mm) in linear gradient mode, at a rate of 0.4 ml/minute. Mobile phase composition: 0.1% aqueous formic acid solution and 0.1% formic acid solution in acetonitrile. Detection of vardenafil and sildenafil in samples was performed using a single quadrupole mass spectrometer with electrospray ionization. Mass spectral analysis was performed in the m/z range: 50–500 with the fragmentation of 50 m/z under positive ionization. The linear dependence for vardenafil was in the range of 7–500 ng/ml; limit of detection and limit of quantification were 5 and 7 ng/ml, respectively. The developed method was tested on specimens of rat urine taking vardenafil at a dose of 0.28 mg/kg. 11.27–13.60 ng of vardenafil were detected in 1 ml of daily urine of the animals. This method is recommended for use in toxicological practice.


Keyword:     Vardenafil HPLC mass spectrometry urine.


Citation:

Osypchuk L, Halkevych I, Davydovych S, Bidnychenko Y. Validation of an HPLC-MS method for the determinatin of vardenafil in rat urine. J Appl Pharm Sci, 2019; 9(08): 079–085.

Copyright: © The Author(s). This is an open-access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

HTML Full Text

INTRODUCTION

Vardenafil (2-{2-ethoxy-5-[(4-ethylpiperazin-1-yl)sulfonyl]phenyl}-5-methyl-7-propyl-1H,4H-imidazo[4,3-f][1,2,4]triazin-4-one) is a representative of the class of phosphodiesterase type 5 inhibitors (PDE5). In medical practice, it is used to treat erectile dysfunction (ED) (Bruzziches et al., 2013; Eve, 2009) and undergoes clinical trials for the treatment of pulmonary arterial hypertension (Jing et al., 2011; Karasu-Minareci et al., 2012).

PDE5 inhibitors, including vardenafil, are a part of the dietary supplements (DS) used to treat ED (Al-Tahami, 2014; Gilard, 2015; Reeuwijk et al., 2013; Singh et al., 2009; Walker et al., 2016). The availability of different DS in the pharmaceutical market, the ability of patients to purchase these medications in online stores, and the lack of knowledge about the presence of PDE5 inhibitors in the DS may have negative effects, especially in case there are contraindications to their use (Chiang et al., 2017; Jackson et al., 2011; Schnetzler et al., 2010).

In medical practice, PDE5 inhibitors should not be combined with nitrovasodilators and inhalation forms of alkyl nitrites—amyl nitrite, butyl nitrite, and isobutyl nitrite (DeBusk et al., 2000; Kostis et al.,2005). Due to the potentiation of the vasodilating effect, the combined use of drugs from these two pharmacological groups leads to an unpredictable decrease in blood pressure, the development of critical uncontrolled hypotension and sudden death (European Medicines Agency, 2010; Kloner RA, 2014; Kloner АR, Henderson L., 2013; Katlowitz and De Rosalio, 2007).

The concentration of vardenafil in blood and plasma is quite high after taking this drug. The literature describes a number of analytical techniques for drug control in these biological fluids. In particular, analytical techniques using fluorimetry (Abu El-Enin et al., 2016), high-performance liquid chromatography (HPLC) with fluorescence detection are described (Cheng et al., 2007), HPLC-UV (Carlucci et al., 2009; Xiao et al., 2013), HPLC- mass spectrometry (MS) (Johnson et al., 2007); LC-MS/MS (Lakea et al., 2010; Rust et al., 2012), UPLC-MS/MS (Proença et al., 2013), and GC-MS (Papoutsis et al., 2011).

According to literature, 2%–6% of the administered dose of vardenafil is excreted with the urine (Bayer Inc, 2015). Therefore, the concentration of vardenafil in this biological matrix is low. Few techniques for the determination of vardenafil and its metabolites in urine have been described. In particular, the determination of vardenafil and its metabolite, desethyl vardenafil, was performed by GC-MS. Extraction from the urine was carried out by liquid extraction, and the extracted compounds were converted to trimethylsilyl derivatives (Strano-Rossi et al., 2010).

There is also a study to detect vardenafil and desethyl vardenafil in urine and blood of volunteers by LC-MS/MS after purification of biological samples using SPE Oasis HLB cartridges (Hasegawa et al., 2012).

Therefore, the purpose of the study was to develop a bioanalytical method for the quantitative determination of vardenafil in rat urine by the HPLC-MS in the presence of metabolites, which could be used for toxicological studies and in the practice of forensic chemical analysis of this biological matrix.


MATERIALS AND METHODS

Chemicals and reagents

Vardenafil hydrochloride trihydrate [≥98% (HPLC), Sigma-Aldrich, USA] and sildenafil citrate—internal standard (Sigma-Aldrich, USA) were used to prepare a series of standard and working solutions.

Nearly, 5 mg of Levitra tablet, Bayer Pharma, were used to prepare solutions administered orally into rats.

Methanol, acetonitrile, orthophosphate acid, and formic acid were of analytical “HPLC” grade (≥99.9%, Sigma-Aldrich, USA). 1,2-dichloroethane and sodium hydroxide were of analytical grade (Merck). The bidistilled water was purified using a Milli-Q purification system (Millipore; Vienna, Austria).

Equipment

Vortex MSV-3500, (Biosan, Latvia) was used for mixing solutions.

Evaporation of organic solvents was performed at 40°C using a TurboVap® Classic LV evaporator (Biotage, Sweden).

Centrifugation was performed on a Sigma 3-16KL refrigerated centrifuge.

The pH control was performed using PH-200 pH-meter (HM Digital, USA) with automatic temperature compensation.

Solid-phase extraction was performed using Oasis HLB cartridges 30 mg (Waters, USA).

Chromatographic conditions

Experimental studies were carried out on the Agilent 1260 Infinity HPLC System liquid chromatograph (Agilent Technologies, USA) with OpenLAB CDS Software.

Chromatographic separation was carried out on a Zorbax SB-C18 reversed-phase column (50 mm × 2.1 mm, 1.8 μm). Composition of the mobile phase: 0.1% aqueous formic acid solution (solution A), 0.1% solution of formate acid in acetonitrile (solution B). The ratios between the volume of A and B solutions in the mobile phase during the analysis were as follows: from the start till 10 minutes—95:5; from 10 to 15 minutes—0:100; 15 minutes—100% of B solution. The mobile phase was given in linear gradient mode, at a rate of 0.4 ml/minute. The column temperature was 40°C. Duration of the analysis was 25 minutes.

Detection of vardenafil and sildenafil in samples was performed using Agilent 6120 single quadrupole mass spectrometer with electrospray ionization.

Mass spectral analysis was performed in the m/z range: 50–500 with the fragmentation of 50 m/z under positive ionization. Ion source for mass spectrometric detection: API-ES. The temperature of nitrogen (gas drier) was 300°C; pressure on a nebulizer was 40 psi, speed of nitrogen in an ion source—10 l/minute.

Animal experiments

Two groups of animals were selected for the study: control and study [10 rats each (males) weighing 190–200 g]. The study and control group of animals had free access to food and water. The rats’ urine from both groups was collected for 24 hours.

The study group of animals was administered an aqueous suspension of “Levitra” for 0.28 mg/kg once. The aqueous suspension was obtained by dissolving the Levitra tablets (5 mg) in 10 ml of water. Nearly, 5 ml of urine were taken for one study.

The rats were kept at the Vivarium Center of Danylo Halytsky Lviv National Medical University. Experiments adhered to ethical standards and approved by the ethical committee of mentioned university (Approval File No. 1/2015).

Animal urine was analyzed immediately after collection or stored in a frozen state until used.

Preparations of standard and sample solutions

Standard solutions of vardenafil (1 mg/ml) and sildenafil (internal standard, 1 mg/ml) were prepared by dissolving the accurately weighed quantity of standard substances in methanol (solutions A). Obtained standard solutions can be used within 3 months, keeping them at 4°C in a dark place.

Methanol solutions of vardenafil and sildenafil containing 10 μg/ml (solution B) were prepared from standard solutions A. A solution of vardenafil containing 1 μg/ml (solution C) was prepared by diluting the solution B in methanol.

For analysis, three parallel series of urine samples with a concentration of vardenafil from 7 to 500 ng/ml (7.0, 50.0, 100.0, 200.0, 300, 400, and 500.0 ng/ml) was prepared. To do this, 100 μl of internal standard (solution B) and 35.0 and 250.0 μl of a solution of vardenafil at a concentration of 1 μg/ml (solution C) and 50.0, 100.0, 150.0, 200.0, and 250 μl of vardenafil solution at a concentration of 10 μg/ml (solution B) were added to 3 ml of pure rat urine. The volumes of biological fluid were adjusted to 5 ml with urine. Samples were stirred for 3 minutes in an orbital shaker and then incubated at 37°C for 60 minutes.

The obtained solutions were used to construct calibration graphs using the internal standard method in the concentration range of 7–500 ng/ml. Similarly, urine specimens were prepared for validation of the method, in which the vardenafil concentration was 7, 200, and 500 ng/ml, and each was subjected to 100 μl of the internal standard. Model samples and validation solutions were subjected to further sampling according to the scheme described below.

Sample preparation

The urine samples were adjusted to pH 7.5 with 30% sodium hydroxide solution, followed by double extraction with 1,2-dichloroethane (5 ml), the duration of single extraction was 10 minutes. The combined dichloroethane fractions were evaporated to dryness. The dry residue was dissolved in 0.4 ml of methanol and diluted to 2 ml with water. The entire volume of the methanol-water solution was purified through SPE cartridges. The cartridges were conditioned with 1 ml of methanol and 1 ml of distilled water. After the test was added, the columns were washed with 2 ml of a universal buffer solution (pH 7.4) and 1 ml of distilled water. Within 5 minutes, the sorbent was dried in a stream of nitrogen; elution was performed with 2 ml of methanol. The speed of all liquids passing through the sorbent was 1 ml/minute. The resulting eluate was evaporated to dryness in a stream of nitrogen and dissolved in 500 μl of methanol. Nearly, 10 μl methanol solution was injected into the chromatograph.

Control urine samples were examined in parallel according to the same mode.


METHOD VALIDATION

Validation of the elaborated technique for the quantitative determination of vardenafil in urine was carried out in accordance with the FDA and EMA guidelines on validation of bioanalytical methods (European Medicines Agency, 2011; Food and Drug Administration, 2013).

Specificity

To prove the specificity of the technique, the chromatograms of control samples of rat urine and those added to the internal standard and working solutions of vardenafil were compared. The concentration of the working solution of vardenafil corresponded to the lower range of the quantitative determination—7 ng/ml, and sildenafil—200 ng/ml.

Linearity and range

The linearity of the method was studied on model urine mixtures in the range of concentrations of vardenafil 7–500 ng/ml.

Applying the least squares method, the equation of the line, which describes the relationship between the ratio of peak areas of vardenafil to the internal standard (sildenafil) and the concentration of vardenafil in urine, ng/ml, was calculated. The limit of detection (LOD) was determined on five model urine samples with a signal-to-noise ratio of at least 3:1. The limit of quantification (LOQ) was defined as the lowest concentration of vardenafil, which can be accurately measured [coefficient of variation (CV) of less than 20%]. The study was conducted on five model urine samples containing 7 ng/ml vardenafil and 200 ng/ml sildenafil (internal standard), independent of the calibration curve.

Figure 1. Chromatogram of drug-free urine samples.

[Click here to view]

Matrix effects

The matrix effect is a direct or non-direct change in the instrument signal due to the presence of other analytes in the sample. In the study of this criterion, six series of biological matrices (urine) were used. The influence of the biological matrix on the ionization efficiency of the analyte and the internal standard was calculated by analyzing the ratio of the peak area of the analyte in the biological sample to the peak area of the analyte in the methanol solution. Normalized by the internal standard, the matrix effect was determined as the ratio of the matrix effect of the analyte to the matrix effect of the internal standard. The matrix effect was studied on the lower (7 ng/ml), middle (200 ng/ml), and upper (500 ng/ml) levels of vardenafil concentrations.

Accuracy and precision

To determine the accuracy and precision of urine samples, the amounts of vardenafil were given, which corresponded to the lower range of quantitative analysis by this method (7 ng/ml), middle (200 ng/ml), and upper level of concentration (500 ng/ml), as well as 200 ng/ml of sildenafil (internal standard). Six samples were prepared with each concentration level.

Concentrations of medications in samples were calculated by the equation of calibration graphs; afterward, the obtained values of concentrations were compared with their nominal value.

Samples were analyzed on the day of preparation and 24 hours after (samples were stored at 4°C).

The degree of extraction of vardenafil from the urine was determined by calculating the ratio of the amount of vardenafil extracted from the biological sample to the amount introduced into the dry residue of the control sample. The same concentration levels were used to determine the accuracy and precision; six samples for each level were prepared.

The short-term and long-term stability of vardenafil, as well as the stability of vardenafil during three freeze-defrost cycles, were studied on model urine samples containing 7 and 500 ng/ml of the medication.


RESULTS

Identification of vardenafil on chromatograms was carried out in absolute retention time, which was 5.820 ± 0.085 minutes under the described conditions of the analysis and the mass spectrum. The retention time of the internal standard (sildenafil) was 6.103 ± 0.052 minutes.

Urine sample chromatogram without medications, as well as urine sample chromatogram with vardenafil concentration of 300 ng/ml and the internal standard (sildenafil) of 200 ng/ml, are shown in Figures 1 and 2. The mass spectra of vardenafil and sildenafil are shown in Figures 4 and 5.

Figure 2. Chromatogram of model urine samples, containing 300 ng/ml vardenafil (1) and 200 ng/ml internal standard (sildenafil) (2).

[Click here to view]

Figure 3. A representative chromatogram of urine samples of rats treated Levitra, Bayer Pharma (1—vardenafil, 2—internal standard (sildenafil)).

[Click here to view]

Figure 4. Mass spectrum of vardenafil in SCAN mode 50–500 m/z (fragmentation 50 m/z).

[Click here to view]

Figure 5. Mass spectrum of sildenafil in SCAN mode 50–500 m/z (fragmentation 50 m/z).

[Click here to view]

Figure 6. Mass spectra of vardenafil extracted from rats urine: I—2-[2-ethoxy-5-(1-hydroxyethyl-amino)sulfonyl]phenyl}-5-methyl-7-propyl-1H,4H-imidazo[4,3-f][1,2,4]triazin-4-one (tR = 4.722 minutes), II—desethyl vardenafil (tR = 5.144 minutes), III—desmethyl vardenafil. (tR = 5.448 minutes).

[Click here to view]

The proposed method for the extraction and determination of vardenafil was used to analyze urine specimens of laboratory rats taking the study medication.

At the same time, the results obtained after the analysis of model urine samples containing vardenafil (Fig. 2) biological fluid samples of rats taking Levitra (Fig. 3) were compared with samples of pure rat urine (Fig. 1). Major metabolites were noted on the chromatograms of urine samples of rats taking Levitra, the peaks of which deviated and were absent on the chromatograms of control urine samples. Their identification was carried out with the obtained mass spectra.

The retention time of the metabolites was 4.72 minutes—degradation product of the piperazine cycle: 2-[2-ethoxy-5-(1-hydroxyethyl-amino)sulfonyl]phenyl}-5-methyl-7-propyl-1H,4H-imidazo[4,3-f][1,2,4]triazin-4-one, 5.14 minutes—desethyl vardenafil and 5.45 minutes—desmethyl vardenafil. Mass spectra of extracted metabolites are shown in Figure 6.

From 11.27 ± 0.28 to 13.60 ± 0.30 ng of vardenafil in 1 ml of daily urine of animals were determined by the proposed method. The total metabolites were about 50% of the concentration level of the medication.

Specificity

In the analysis of urine specimens, chromatograms do not show peaks with the retention time corresponding to vardenafil, its metabolites, and sildenafil.

The mass spectrum of vardenafil extracted from the urine corresponds to the mass spectrum of the standard sample of vardenafil and it is characterized by such signals of ions—489, 491 m/z.

Linearity and range

Quantitative analysis was carried out by the internal standard method. Applying the least squares method, the equation of the line, which describes the relationship between the ratio of peak areas of vardenafil to the internal standard and the concentration of vardenafil in urine, ng/ml, was calculated. The linear dependence for vardenafil was in the range of 7–500 ng/ml. The equation of the calibration graph in the concentration ranges from 7 to 500 ng/ml is described by the dependence Y = 0.0038 X − 0.0116, where: Y—the ratio between peak areas of vardenafil and internal standard, X—concentration of vardenafil, ng/ml.

The LOD of vardenafil by HPLC/MS on chromatograms was determined by the signal-to-noise ratio of 3:1 and the LOQ by the ratio of 5:1. The LOD of vardenafil was 5 ng/ml and the LOQ was 7 ng/ml.

Accuracy and precision

The results of the precision and accuracy of the evaluation of vardenafil in urine samples by the HPLC-MS are presented in Table 1, and in the urine of the rats taking Levitra tablets in Table 2. The developed method is accurate and reproducible since the data are repeatable within one day and on different days.

Matrix effects

The matrix effect, normalized by the internal standard for six samples of biological fluid, is on average 1.14; the average absolute deviation of the matrix effect is 2.54. According to the recommendations on validation, the CV of the matrix effect, normalized by the internal standard for six samples of urine, should not exceed 15%. The obtained results meet the established criteria and indicate the absence of the influence of endogenous components on the results of the quantitative determination of vardenafil in the biological matrices under study. The degree of extraction of vardenafil from urine and normalized matrix effects (n = 6) are given in Table. 3

Stability

Vardenafil was stable in model urine samples stored at room temperature for 24 hours. Three freeze-defrost cycles of urine samples with vardenafil also demonstrated the stability of the medication. The deviation also did not exceed the permissible standards with prolonged storage of urine samples with vardenafil (within 60 days). The results of the studies are presented in Table 4.


DISCUSSION

The aim of the study was to develop a bioanalytical method for the quantitative determination of vardenafil in the urine in the presence of metabolites by the HPLC-MS, which could be used for toxicological studies and in the practice of forensic chemical analysis. The limit for detecting vardenafil in urine was 5 ng/ml, and the limit of quantitative determination was 7 ng/ml, with a sampling of 5 ml of urine per study. The proposed method determined 11.27–13.60 ng of vardenafil in 1 ml of daily animal urine. The total content of metabolites was about 50% of the drug concentration level. The proposed evaluation method allows determining vardenafil in urine taken at the therapeutic dose.

Table 1. Accuracy and precision of vardenafil determination method in model urine samples of rats.

[Click here to view]

Table 2. Accuracy and precision of vardenafil determination method in urine of Levitra treated rats.

[Click here to view]

Table 3. The degree of extraction and normalized matrix effects in vardenafil determination in urine samples (n = 6).

[Click here to view]

Table 4. Stability of vardenafil in urine.

[Click here to view]


CONCLUSION

A simple and reliable method for the determination of vardenafil in the urine of white rats is described using HPLC-MS on the Zorbax SB-C18 reversed-phase column. The proposed conditions of the chromatographic analysis provide for the rapid determination of vardenafil and its metabolites, as well as high selectivity, accuracy, and reproducibility of the results, lack of influence of the matrix effect. The results of the validation of the method showed its suitability for chemical and toxicological analysis.


FINANCIAL SUPPORT AND SPONSORSHIP

None.


CONFLICT OF INTERESTS

The authors declare no conflict of interest.


REFERENCES

Abu El-Enin MA, Al-Ghaffar HMS, El-Sherbiny DT, El-Wasseef DR, El-Ashry SM. Validated spectrofluorimetric method for determination of two phosphodiesterase inhibitors tadalafil and vardenafil in pharmaceutical preparations and spiked human plasma. Luminescence, 2016; 31:173–8.CrossRef

Al-Tahami K. Determination of sildenafil, vardenafil and tadalafil in dietary supplements sold in the Yemeni market. Int J Sci Res, 2014; 3(4):403–5. CrossRef

Bayer Inc. LEVITRA® (vardenafil tablets 5 mg, 10 mg, 20 mg of vardenafil, as vardenafil hydrochloride)—Product Monograph. Mississauga, Ontario, 2015. Available via https://omr.bayer.ca/omr/online/levitra-pm-en.pdf (Accessed 9 March 2019).

Bruzziches R, Francomano D, Gareri P, Lenzi A, Aversa A. An update on pharmacological treatment of erectile dysfunction with phosphodiesterase type-5 inhibitors. Expert Opin Pharmacother, 2013; 14(10):1333–44. CrossRef

Carlucci G, Palumbo P, Iuliani P, Palumbo G. Development of a method for the determination of vardenafil in human plasma by high performance liquid chromatography with UV detection. Biomed Chromatogr, 2009; 23(7):759–63. CrossRef

Cheng CL, Kang GJ, Chou CH. Development and validation of a high-performance liquid chromatographic method using fluorescence detection for the determination of vardenafil in small volumes of rat plasma and bile. J Chromatogr A, 2007; 1154.1.2:222–9. CrossRef

Chiang J, Yafi FA, Dorsey PJ, Hellstrom WJ. The dangers of sexual enhancement supplements and counterfeit drugs to "treat" erectile dysfunction. Transl Androl Urol, 2017; 6(1):12–9. CrossRef

DeBusk R, Drory Y, Goldstein I, Jackson G, Kaul S, Kimmel SE, Kostis JB, Kloner RA, Lakin M, Meston CM, Mittleman M, Muller JE, Padma-Nathan H, Rosen RC, Stein RA, Zusman R. Management of sexual dysfunction in patients with cardiovascular disease: recommendations of the Princeton Consensus Panel. Am J Cardiol, 2000; 86(2):175–81. CrossRef

European Medicines Agency. Levitra (International Nonproprietary Name: vardenafil). CHMP assessment report, London, 2010. Available via https://www.ema.europa.eu/en/documents/variation-report/levitra-h-c-475-x-28-epar-assessment-report-extension_en.pdf (Accessed 10 March 2019).

European Medicines Agency (EMEA/CHMP/EWP/192217/2009). Guideline on bioanalytical method validation, London, 2011.

Eve D. Drug profile: Levitra. Medicine discovery, 2009. Available via http://www.discoverymedicine.com/David-Eve/2009/05/23/drug-profile-levitrа (Accessed 15 March 2019).

Food, Drug Administration Centre for Drug Evaluation and Research (FDA). Guidance for industry-bioanalytical method validation, center for drug evaluation and research. US Department for Health and Human Services, 2013, Silver Spring, MD, 2001.

Gilard V, Balayssac S, Tinaugus A, Martins N, Martino R, Malet-Martino M. Detection, identification and quantification by 1H NMR of adulterants in 150 herbal dietary supplements marketed for improving sexual performance. J Pharm Biomed Anal, 2015; 102:476–93. CrossRef

Hasegawa K, Suzuki O, Gonmori K, Yamagishi I, Nozawa H, Watanabe K. Simultaneous analysis of sildenafil, vardenafil, tadalafil, and their desalkyl metabolites in human whole blood and urine by isotope dilution LC-MS-MS. Forensic Toxicol, 2012; 30:25–32. CrossRef

Jackson G, Arver S, Banks I,Stecher V.J. Counterfeit phosphodiesterase type 5 inhibitors pose significant safety risks. Int J Clin Pract, 2011; 64.4:497–504. CrossRef

Jing ZC, Yu ZX, Shen JY, Wu BX, Xu KF, Zhu XY, Pan L, Zhang ZL, Liu XQ, Zhang YS, Jiang X, Galiè N. Vardenafil in pulmonary arterial hypertension: a randomized, double-blind, placebo-controlled study. Am J Respir Crit Care Med, 2011; 183(12):1723–9. CrossRef

Johnson RD, Lewis RJ, Angier MK. The postmortem distribution of vardenafil (Levitra) in an aviation accident victim with an unusually high blood concentration. J Anal Toxicol, 2007; 31(6):328–33. CrossRef

Karasu-Minareci E, Ozbudak I.H, Ozbilim G, Sadan G. Acute effects of vardenafil on pulmonary artery responsiveness in pulmonary hypertension. Sci World J, 2012; 2012:718279; doi:10.1100/2012/718279 CrossRef

Katlowitz NM, De Rosalio A. Phosphodiesterase type 5 inhibitors and nitrates: is concomitant use safe? Are there alternatives? Curr Sexual Health Rep, 2007; 4(2):59–63. CrossRef

Kloner RA. Novel phosphodiesterase type 5 inhibitors: assessing hemodynamic effects and safety parameters. Clin Cardiol, 2004; 27(1):20–5. CrossRef

Kloner АR, Henderson L. Sexual function in patients with chronic angina pectoris. Am J Cardiol, 2013; 111(1):1671–6. CrossRef

Kostis JB, Jackson G, Rosen R, Barrett-Connor E, Billups K, Burnett A.L, Carson G, Cheitlin M, Debusk M, Fonseca V, Ganz P, Goldstein I, Guay A, Hatzichristou D, Hollander J. E, Hutter A, Katz S, Kloner R.A, Mittleman M, Montorsi F, Montorsi P, Nehra A, Sadovsky R, Shabsigh R. Sexual dysfunction and cardiac risk (the Second Princeton Consensus Conference). Am J Cardiol, 2005; 96(2):313–21. CrossRef

Lakea ST, Altmanb PM, Vaismanc J, Addisona RS. Validated LC-MS/MS assay for the quantitativ determination of vardenafi l in human plasma and its application to a pharmacokinetic study. Biomed Chromatogr, 2010; 24:846–51; doi:10.1002/bmc.1375 CrossRef

Papoutsis I, Nikolaou P, Athanaselis S, Pistos C, Maravelias C, Spiliopoulou C. A fully validated method for the determination of vardenafil in blood using gas chromatography/mass spectrometry. J Mass Spectrom, 2011;46(1):71–6. CrossRef

Proença P, Mustra C, Marcos M, Franco J.M, Corte-Real F, Vieira D.N. Validated UPLC-MS/MS assay for the determination of synthetic phosphodiesterase type-5 inhibitors in postmortem blood samples. J Forensic Leg Med, 2013; 20(6):655–8. CrossRef

Reeuwijk NM, Venhuis BJ, de Kaste D, Hoogenboom LA, Rietjens IM, Martena MJ. Sildenafil and analogous phosphodiesterase type 5 (PDE-5) inhibitors in herbal food supplements sampled on the Dutch market. Food Addit Contam Part A Chem Anal Control Expo Risk Assess, 2013; 30(12):2027–34. CrossRef

Rust KY, Wilkens H, Kaiser R, Bregel D, Wilske J, Kraemer T. Detection and validated quantification of the phosphodiesterase type 5 inhibitors sildenafil, vardenafil, tadalafil, and 2 of their metabolites in human blood plasma by LC-MS/MS-application to forensic and therapeutic drug monitoring cases. Ther Drug Monit, 2012; 34(6):729–35. CrossRef

Schnetzler G, Banks I, Kirby M, Zou KH, Symonds T. Characteristics, behaviors, and attitudes of men bypassing the healthcare system when obtaining phosphodiesterase type 5 inhibitors. J Sex Med, 2010; 7(3):1237–46. CrossRef

Singh S, Prasad B, Savaliya AA, Shah RP, Gohil VM, Kaur A. Strategies for characterizing sildenafil, vardenafil, tadalafil and their analogues in herbal dietary supplements, and detecting counterfeit products containing these drugs. TrAC Trends Analyt Chem, 2009; 28(1):13–28. CrossRef

Strano-Rossi S, Anzillotti L, Torre X, Botre F. A gas chromatography/mass spectrometry method for the determination of sildenafil, vardenafil and tadalafil and their metabolites in human urine. Rapid Commun Mass Spectrom, 2010; 24:1697–706. CrossRef

Walker MJ, Naughton DP, Deshmukh N, Burns, DT. A review of methods for the simultaneous detection of illegal ingredients in food supplements. J Assoc Public Anal, 2016; 44:51–6. Available via http://www.apajournal.org.uk/2016_0051-0066.pdf (Accessed 20 February 2019).

Xiao C, Tang M, Li J, Yin CR, Xiang G, Xu L. Determination of sildenafil, vardenafil and aildenafil in human plasma by dispersive liquid–liquid microextraction-back extraction based on ionic liquid and high performance liquid chromatography-ultraviolet detection. J Chromatogr B Anal Technol Biomed Life Sci, 2013; 931:111–6.CrossRef

Reference

Abu El-Enin MA, Al-Ghaffar HMS, El-Sherbiny DT, El- Wasseef DR, El-Ashry SM. Validated spectrofluorimetric method for determination of two phosphodiesterase inhibitors tadalafil and vardenafil in pharmaceutical preparations and spiked human plasma. Luminescence, 2016; 31:173-8. https://doi.org/10.1002/bio.2941

Al-Tahami K. Determination of sildenafil, vardenafil and tadalafil in dietary supplements sold in the Yemeni market. Int J Sci Res, 2014; 3(4):403-5. https://doi.org/10.15373/22778179/APR2014/143

Bayer Inc. LEVITRA® (vardenafil tablets 5 mg, 10 mg, 20 mg of vardenafil, as vardenafil hydrochloride)-Product Monograph. Mississauga, Ontario, 2015. Available via https://omr.bayer.ca/omr/online/ levitra-pm-en.pdf (Accessed 9 March 2019).

Bruzziches R, Francomano D, Gareri P, Lenzi A, Aversa A. An update on pharmacological treatment of erectile dysfunction with phosphodiesterase type-5 inhibitors. Expert Opin Pharmacother, 2013; 14(10):1333-44. https://doi.org/10.1517/14656566.2013.799665

Carlucci G, Palumbo P, Iuliani P, Palumbo G. Development of a method for the determination of vardenafil in human plasma by high performance liquid chromatography with UV detection. Biomed Chromatogr, 2009; 23(7):759-63. https://doi.org/10.1002/bmc.1181

Cheng CL, Kang GJ, Chou CH. Development and validation of a high-performance liquid chromatographic method using fluorescence detection for the determination of vardenafil in small volumes of rat plasma and bile. J Chromatogr A, 2007; 1154.1.2:222-9. https://doi.org/10.1016/j.chroma.2007.03.077

Chiang J, Yafi FA, Dorsey PJ, Hellstrom WJ. The dangers of sexual enhancement supplements and counterfeit drugs to "treat" erectile dysfunction. Transl Androl Urol, 2017; 6(1):12-9. https://doi.org/10.21037/tau.2016.10.04

DeBusk R, Drory Y, Goldstein I, Jackson G, Kaul S, Kimmel SE, Kostis JB, Kloner RA, Lakin M, Meston CM, Mittleman M, Muller JE, Padma-Nathan H, Rosen RC, Stein RA, Zusman R. Management of sexual dysfunction in patients with cardiovascular disease: recommendations of the Princeton Consensus Panel. Am J Cardiol, 2000; 86(2):175-81. https://doi.org/10.1016/S0002-9149(00)00896-1

European Medicines Agency. Levitra (International Nonproprietary Name: vardenafil). CHMP assessment report, London, 2010. Available via https://www.ema.europa.eu/en/documents/variation-report/levitra-h-c-475-x-28-epar-assessment-report-extension_en.pdf (Accessed 10 March 2019).

European Medicines Agency (EMEA/CHMP/ EWP/192217/2009). Guideline on bioanalytical method validation, London, 2011.

Eve D. Drug profile: Levitra. Medicine discovery, 2009. Available via http://www.discoverymedicine.com/David-Eve/2009/05/23/ drug-profile-levitrа (Accessed 15 March 2019).

Food, Drug Administration Centre for Drug Evaluation and Research (FDA). Guidance for industry-bioanalytical method validation, center for drug evaluation and research. US Department for Health and Human Services, 2013, Silver Spring, MD, 2001.

Gilard V, Balayssac S, Tinaugus A, Martins N, Martino R, Malet-Martino M. Detection, identification and quantification by 1H NMR of adulterants in 150 herbal dietary supplements marketed for improving sexual performance. J Pharm Biomed Anal, 2015; 102:476-93. https://doi.org/10.1016/j.jpba.2014.10.011

Hasegawa K, Suzuki O, Gonmori K, Yamagishi I, Nozawa H, Watanabe K. Simultaneous analysis of sildenafil, vardenafil, tadalafil, and their desalkyl metabolites in human whole blood and urine by isotope dilution LC-MS-MS. Forensic Toxicol, 2012; 30:25-32. https://doi.org/10.1007/s11419-011-0125-2

Jackson G, Arver S, Banks I,Stecher V.J. Counterfeit phosphodiesterase type 5 inhibitors pose significant safety risks. Int J Clin Pract, 2011; 64.4:497-504. https://doi.org/10.1111/j.1742-1241.2009.02328.x

Jing ZC, Yu ZX, Shen JY, Wu BX, Xu KF, Zhu XY, Pan L, Zhang ZL, Liu XQ, Zhang YS, Jiang X, Galiè N. Vardenafil in pulmonary arterial hypertension: a randomized, double-blind, placebo-controlled study. Am J Respir Crit Care Med, 2011; 183(12):1723-9. https://doi.org/10.1164/rccm.201101-0093OC

Johnson RD, Lewis RJ, Angier MK. The postmortem distribution of vardenafil (Levitra) in an aviation accident victim with an unusually high blood concentration. J Anal Toxicol, 2007; 31(6):328-33. https://doi.org/10.1093/jat/31.6.328

Karasu-Minareci E, Ozbudak I.H, Ozbilim G, Sadan G. Acute effects of vardenafil on pulmonary artery responsiveness in pulmonary hypertension. Sci World J, 2012; 2012:718279; doi:10.1100/2012/718279 https://doi.org/10.1100/2012/718279

Katlowitz NM, De Rosalio A. Phosphodiesterase type 5 inhibitors and nitrates: is concomitant use safe? Are there alternatives? Curr Sexual Health Rep, 2007; 4(2):59-63. https://doi.org/10.1007/s11930-007-0003-4

Kloner RA. Novel phosphodiesterase type 5 inhibitors: assessing hemodynamic effects and safety parameters. Clin Cardiol, 2004; 27(1):20-5. https://doi.org/10.1002/clc.4960271306

Kloner АR, Henderson L. Sexual function in patients with chronic angina pectoris. Am J Cardiol, 2013; 111(1):1671-6. https://doi.org/10.1016/j.amjcard.2013.02.009

Kostis JB, Jackson G, Rosen R, Barrett-Connor E, Billups K, Burnett A.L, Carson G, Cheitlin M, Debusk M, Fonseca V, Ganz P, Goldstein I, Guay A, Hatzichristou D, Hollander J. E, Hutter A, Katz S, Kloner R.A, Mittleman M, Montorsi F, Montorsi P, Nehra A, Sadovsky R, Shabsigh R. Sexual dysfunction and cardiac risk (the Second Princeton Consensus Conference). Am J Cardiol, 2005; 96(2):313-21. https://doi.org/10.1016/j.amjcard.2005.03.065

Lakea ST, Altmanb PM, Vaismanc J, Addisona RS. Validated LC-MS/MS assay for the quantitativ determination of vardenafi l in human plasma and its application to a pharmacokinetic study. Biomed Chromatogr, 2010; 24:846-51; doi:10.1002/bmc.1375 https://doi.org/10.1002/bmc.1375

Papoutsis I, Nikolaou P, Athanaselis S, Pistos C, Maravelias C, Spiliopoulou C. A fully validated method for the determination of vardenafil in blood using gas chromatography/mass spectrometry. J Mass Spectrom, 2011;46(1):71-6. https://doi.org/10.1002/jms.1868

Proença P, Mustra C, Marcos M, Franco J.M, Corte-Real F, Vieira D.N. Validated UPLC-MS/MS assay for the determination of synthetic phosphodiesterase type-5 inhibitors in postmortem blood samples. J Forensic Leg Med, 2013; 20(6):655-8. https://doi.org/10.1016/j.jflm.2013.03.002

Reeuwijk NM, Venhuis BJ, de Kaste D, Hoogenboom LA, Rietjens IM, Martena MJ. Sildenafil and analogous phosphodiesterase type 5 (PDE-5) inhibitors in herbal food supplements sampled on the Dutch market. Food Addit Contam Part A Chem Anal Control Expo Risk Assess, 2013; 30(12):2027-34. https://doi.org/10.1080/19440049.2013.848294

Rust KY, Wilkens H, Kaiser R, Bregel D, Wilske J, Kraemer T. Detection and validated quantification of the phosphodiesterase type 5 inhibitors sildenafil, vardenafil, tadalafil, and 2 of their metabolites in human blood plasma by LC-MS/MS-application to forensic and therapeutic drug monitoring cases. Ther Drug Monit, 2012; 34(6):729-35. https://doi.org/10.1097/FTD.0b013e31827318b8

Schnetzler G, Banks I, Kirby M, Zou KH, Symonds T. Characteristics, behaviors, and attitudes of men bypassing the healthcare system when obtaining phosphodiesterase type 5 inhibitors. J Sex Med, 2010; 7(3):1237-46. https://doi.org/10.1111/j.1743-6109.2009.01674.x

Singh S, Prasad B, Savaliya AA, Shah RP, Gohil VM, Kaur A. Strategies for characterizing sildenafil, vardenafil, tadalafil and their analogues in herbal dietary supplements, and detecting counterfeit products containing these drugs. TrAC Trends Analyt Chem, 2009; 28(1):13-28. https://doi.org/10.1016/j.trac.2008.09.004

Strano-Rossi S, Anzillotti L, Torre X, Botre F. A gas chromatography/mass spectrometry method for the determination of sildenafil, vardenafil and tadalafil and their metabolites in human urine. Rapid Commun Mass Spectrom, 2010; 24:1697-706. https://doi.org/10.1002/rcm.4568

Walker MJ, Naughton DP, Deshmukh N, Burns, DT. A review of methods for the simultaneous detection of illegal ingredients in food supplements. J Assoc Public Anal, 2016; 44:51-6. Available via http://www. apajournal.org.uk/2016_0051-0066.pdf (Accessed 20 February 2019).

Xiao C, Tang M, Li J, Yin CR, Xiang G, Xu L. Determination of sildenafil, vardenafil and aildenafil in human plasma by dispersive liquid- liquid microextraction-back extraction based on ionic liquid and high performance liquid chromatography-ultraviolet detection. J Chromatogr B Anal Technol Biomed Life Sci, 2013; 931:111-6. https://doi.org/10.1016/j.jchromb.2013.05.013

Article Metrics
602 Views 80 Downloads 682 Total

Year

Month

Related Search

By author names